Several run lists are now available in the GDA giving coverage with the detector set to either 160 mm (in practice, this is as close in as possible for a reasonable omega sweep) and further back at 300 mm.
At a detector distance of 160 mm, 2θ=30° is the maximum value where the beamstop is still visible.
At a detector distance of 300 mm, 2θ=20° is the maximum value where the beamstop is still visible.
Run lists do not need to be used in their entirety – just select the rows required to be collected.
The Strategies page should provide some guidance to each run list and can used to modify or prepare your own run list.
The standard sphere can be loaded by clicking on the "Load Default Run List" button
Or, other options can be loaded using the "Select Collection Strategy" button.
This will open a window towards the top left of the screen – Default always shows up as the default option.
NOTE: The run list selected will overwrite any existing runs in the table.
Diamond Light Source is the UK's national synchrotron science facility, located at the Harwell Science and Innovation Campus in Oxfordshire.
Copyright © 2022 Diamond Light Source
Diamond Light Source Ltd
Diamond House
Harwell Science & Innovation Campus
Didcot
Oxfordshire
OX11 0DE
Diamond Light Source® and the Diamond logo are registered trademarks of Diamond Light Source Ltd
Registered in England and Wales at Diamond House, Harwell Science and Innovation Campus, Didcot, Oxfordshire, OX11 0DE, United Kingdom. Company number: 4375679. VAT number: 287 461 957. Economic Operators Registration and Identification (EORI) number: GB287461957003.